Message
Remedial Action
Flow meter data read failure
The data on the flow meter EEPROM could not be read.
Verify that the flow meter is installed properly and perform a self
test.
Flow meter failure
Invalid values are reported for one or more pressure transducers of
the flow meter.
Verify that the flow meter is installed properly.
Flow meter missing
Communication with the flow meter failed.
Verify that the flow meter is installed properly.
Flow meter recently changed.
Please use the zero balance
test and then calibrate both
working pressure
transducers.
After installing the ProFlow flow meter,
Zero Balance Test
and the
CalibrateWP
command have not been performed.
1. Perform the
Zero Balance Test
(
→
page 172).
2. From Chromeleon, perform the
CalibrateWP
command for both
pump heads.
Flow meter recently changed.
Please calibrate (x) working
pressure transducer.
(where x specifies which pressure transducer is affected)
After installing a Classic flow meter, the
Pressure Transducer Test
and
CalibrateWP
command have not been performed.
1. Run the
Pressure Transducer Test
(
→
page 172).
2. From Chromeleon, perform the
CalibrateWP
command for the
pump head.
Flow meter temperature out
of range
The temperature of the flow is outside the allowed range.
The flow meter may be defective. Replace the flow meter
(
→
page 202).
Flow meter data version
mismatch
The firmware installed in the module does not support the full
functionality of the flow meter.
You should update the firmware (
→
page 217).
Flow selector changed.
Please use the pressure
transducer test and perform
calibration.
Calibration has not been performed after the Classic flow meter or
flow selector has been replaced.
Perform calibration by running the
Pressure Transducer Test
(
→
page 172).
Flow selector missing
The flow selector in the Classic flow meter may not be installed or
the installation may not be correct.
Verify that the flow selector is installed properly.
Flow selector version
mismatch
The firmware installed in the module does not support the full
functionality of the flow selector in the Classic flow meter.
You should update the firmware (
→
page 217).
Gas leak detected! Absolute
threshold exceeded.
—or—
Gas leak detected! Relative
threshold exceeded.
—or—
Gas leak detected! Gradient
threshold exceeded.
The gas sensor detected excessive gas in the column chamber.
Eliminate the cause. Tighten leaking connection and vent the column
chamber.
Содержание Dionex UltiMate 3000 Series
Страница 2: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Operating Instructions ...
Страница 8: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page vi Operating Instructions ...
Страница 24: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 16 Operating Instructions ...
Страница 62: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 54 Operating Instructions ...
Страница 92: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 84 Operating Instructions ...
Страница 246: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 238 Operating Instructions ...
Страница 262: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 254 Operating Instructions ...