UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Operating Instructions
Page
193
Wait until the
PistonPositionStatus
property reports
Operational
before you
continue with the next steps.
6.
Flush the pump head with seal wash solution (
→
step 1 on page 186).
7.
To prevent contaminants from entering the HPLC system, thoroughly rinse the pump
(
→
page 72).
8.
Test the pump for leakage (
→
page 211). Tighten leaking connections.
9.
Update the service information in Chromeleon (
→
page 132).