UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Operating Instructions
Page
139
3.
If the module was powere
d off by the power switch for a longer period (≥
1 day),
allow a warm-up period of 1 hour (if a ProFlow flow meter is installed) or 24 hours (if
a Classic flow meter is installed) to achieve optimum reproducibility. You can shorten
the warm-up period if you set the module to standby mode instead of powering it off
completely.
♦
If a Classic flow meter is installed, you can shorten the warm-up period to 1 to
2 hours.
♦
If a ProFlow flow meter is installed, you can shorten the warm-up period to
15 minutes.
4.
Solvents will last several weeks. Replace solvents every 2-4 weeks to avoid large
solvent composition changes or bacterial growth. Replace solvents rather than add up
and time the replacements to fit in between sample sets.
5.
Particularly when using pre-mixed solvents, make sure that the solvents are always
fresh and that they have been properly prepared. After preparation, degas the solvents
by ultrasonication to remove the excess air that could have introduced from the
mixing.
6.
Before
running any calibration routines, perform a purge cycle (
→
page 72) to ensure
that the pump is properly primed.
7.
Flow meter settings:
♦
For Classic flow meters:
Run a viscosity measurement to determine and set the viscosity of
unknown
solvents
or solvent mixture (
→
section 5.5.3.2, page 115). With
known
solvents, select the
known solvent viscosities from the viscosity list (
→
section 4.5.2, page 69) rather
than repeating the test.
Viscosity Measurement functions also as a diagnostics test, providing a
Passed
,
Limited
, or
Failed
result. With
Limited
or
Failed
, follow the instructions provided
by the wizard and repeat the test.
♦
For ProFlow flow meters:
Run a solvent calibration to calibrate the flow meter for
unknown
solvents
(
→
section 5.5.3.1, page 114). With
known
solvents, select the known solvent type
from the flow meter solvent type list (
→
section 4.5.2, page 69) rather than repeating
the calibration procedure.
Содержание Dionex UltiMate 3000 Series
Страница 2: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Operating Instructions ...
Страница 8: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page vi Operating Instructions ...
Страница 24: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 16 Operating Instructions ...
Страница 62: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 54 Operating Instructions ...
Страница 92: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 84 Operating Instructions ...
Страница 246: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 238 Operating Instructions ...
Страница 262: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 254 Operating Instructions ...