UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Operating Instructions
Page
201
•
Recommended
Run the
General Leak Test
from Chromeleon (
→
page 134).
•
In rare cases, new seals may still show an increased leakage rate after several hours of
operation.
If leakage is observed with new piston seals, operate the pump for at least 2 hours at
35 MPa to run in the seals. (It does not help if you tighten the pump head screw more.)
Tip:
If the piston seals do not yet seal sufficiently tight, try the following: Have
the pump build up high pressure (
→
page 211) and hold the pressure for
some minutes.