Safety
3.1 General Safety Guidelines
•
Before using your ARL EQUINOX 5000/5500 X-Ray diffractometer, read this instruction
manual completely and understand it. Please take notice of all warnings, cautions and
notes to ensure the correct and safe functioning of the instrument.
•
Never place the instrument near an open flame or in an excessively hot environment.
The instrument must not be installed in an ATEX explosive hazard area.
•
Never place the instrument in locations where excessive moisture or corrosive materials
are present.
•
The instrument construction provides protection against the risk of electrical shock by
grounding the appropriate metal parts. The protection may not work properly unless the
power cord is connected to a properly grounded socket. It is customer's responsibility to
assure that a proper ground connection is provided.
•
Never use flammable or corrosive fluids with this unit.
•
Never use gases other than described in specifications.
•
Never block the fan grids or the bottom of the instrument to allow a proper ventilation.
•
If the instrument is used in a manner not specified by the Thermo Fisher Scientific,
the protection provided by the instrument may be impaired.
•
Only accessories that comply with Thermo Fisher Scientific specifications are allowed to
be used with ARL EQUINOX 5000/5500 X-Ray diffractometer.
•
Transport of the instrument should be done with extreme care and should be done with
an appropriate lifting system, ideally a forklift truck with forks height not exceeding
110 mm.
•
Never operate a damaged or a leaking instrument.
•
Always turn the instrument OFF and disconnect the power cord from the power source
before performing any service or maintenance tasks, or before moving the instrument.
•
Never operate the instrument with a damaged power cord.
•
Never turn the instrument ON without a proper cooling fluid.
•
Never block the fan outputs.