Chapter |
General Information
Confidential, limited rights
PN AA83865 | Revision A | 25-Oct-2019
Instruction Notice, ARL EQUINOX 5000-5500
Page
15
Water cooling and gas
connections
1/ 2- Gas connection for BTS
furnace
3 /4- Cooling water for BTS
furnace
1 /2- Cooling water for HTK
3 /4- not used for HTK furnace
5- Gas inlet for external cylinder
6- Outlet gas valve
7- Gas outlet from the detector
8/9- Water IN and OUT for
cooling
•
Power: The electrical connection should be done by qualified personnel only.
•
Vacuum: It is recommended to use a peristaltic pump in order to achieve vacuum in the
optic chamber. This will minimize the background noise and preserve the quality of the
layering of the mirror.
•
Gas in: Gas in allows to connect the instrument to an external gas cylinder.
•
Gas out: Gas out is the exhaust of the detector flush. By using the "flow valve" it is
possible to close the flushing and lower the gas consumption. If the host space of the
instrument is too small, it is necessary to connect to an external exhaust.
•
RJ45: The RJ45 connector serves to connect the instrument to a computer using the
TCP/IP protocol.
•
USB: The USB connector serves to connect the instrument to a computer for the
monitoring of the adjustment of the sample holder.
SERVICE!
The service engineer can connect the X-Ray video for beam alignment. The use of
this connector is explained in the Service Guide.