12 Appendix 3: Periodic Preventive Maintenance
Period
Action
Indication
Schedule
Monthly
Check the
detector
calibration
Refer to the SymphoniX user guide
Monthly
Check the
remaining gas
pressure in the
gas cylinder
When
necessary
Replace the
gas cylinder
Contact Thermo Fisher Scientific or an authorized person.
When
necessary
Clean filter of
X-Ray tube
Contact Thermo Fisher Scientific or an authorized person.
Annually
Check the
emergency
stop button
Pushing the emergency stop button must switch off the
instrument. This procedure can be done by the customer,
refer to Section Check Emergency Stop Button on page
Annually
Check the
beam
alignment
Contact Thermo Fisher Scientific or an authorized person.
Annually
Check the gas
circuit tightness
Contact Thermo Fisher Scientific or an authorized person.
Annually
Check the
safety
interlocks
Contact Thermo Fisher Scientific or an authorized person.