Chapter |
General Information
Confidential, limited rights
PN AA83865 | Revision A | 25-Oct-2019
Instruction Notice, ARL EQUINOX 5000-5500
Page
12
4
General Information
4.1 Front View
1 - Door locking system (safety interlock)
2 - Front panel
3 - Security beacons
4 - Ventilation grids
5 - Gas window (reading of remaining gas
cylinder pressure)
6
– Ventilation fans
Height : 1915 mm
Width : 1214 mm
Depth : 828 mm
Weight: 400 kg
The door can be opened by pulling. The door is blocked while the instrument is in operation,
the door lock is released when the shutter is closed.
CAUTION!
Do not avoid the correct functioning of the door interlock. When the shutter closes,
wait 5 seconds so that the door unlocks. Trying to open the door too early when the
door is still locked may block the door. In this case, the door will be released after a
complete instrumental restart: the main breaker located at the back of the instrument
must be set to OFF and ON again in order to restart the safety loop.