Table 3 Variable set when changed
(continued)
Variable
Type
Value/Description
WaitingUserAck
Enum
Instrument has entered pause state. This will
also cause the
ProtocolState
to transition to
ProtocolPaused
.
The initiating event for this pause state is
described:
•
NotWaiting
- Not waiting for User
acknowledge
•
WaitingBubbleAck
- Waiting for User
acknowledge following pause on a bubble
trigger event
•
WaitingPressureAck
- Waiting for User
acknowledge following pause on pressure
trigger event
•
UserPausedButton
– The user paused
the protocol via the pause button on the
instrument
•
UserPausedOPCUA
– The user paused
the protocol via the OPCUA interface
•
In order to acknowledge the event;
The user can press a play/advance buttons (if
enabled) on the instrument.
The MES can write the:
•
ProtocolState
to
RunProtocol
to continue
with the current step
or
•
ProtocolState
to
SkipToNextStep
to
advance
Chapter 4
OPC-UA interface
OPC-UA interface
4
CTS
™
Rotea
™
Counterflow Centrifugation System User Guide
99
Содержание A44769
Страница 131: ......