(continued)
Alarm or
message
Critical
alarm
Warning
Instrument response
GUI message
Centrifuge error
Yes
—
Trigger instrument stop state
Centrifuge error
Protocol run has stopped.
Press the stop button on
the insturment, close manual
clamps and open and close
the instrument door to reset.
Restart the protocol or select
a protocol step to start from. If
error persists, abort the run.
Valve error
Yes
—
Trigger instrument stop state
Valve error
Protocol run has stopped.
Press the stop button on
the instrument, close manual
clamps and open and close
the instrument door to reset.
Restart the protocol or select
a protocol step to start from. If
error persists, abort the run.
Out of balance
condition
Yes
—
Trigger instrument stop state
Instrument out of balance
Protocol run has stopped.
Press the stop button to clear
the error, then restart the
protocol or select a step to
start from. If error persists,
abort the run.
Moisture in bowl
Yes
—
Trigger instrument stop state
Centrifuge leak detected
Protocol run has stopped.
Press the stop button to
clear the error or temporarily
override the moisture sensor if
safe to do so. If error persists,
abort the run.
Chapter 3
Basic instrument operation
Run a protocol
3
68
CTS
™
Rotea
™
Counterflow Centrifugation System User Guide