Instrument properties
Category
Property
Value
Electrical
Supply voltage
100–240 V AC ±10%, 50/60 Hz
Phases
Single
Maximum Rated Input Current
5 A
Fuses
2 x 5 A
Sound level
Maximum sound level
70 dBA (measured 1 m from instrument)
Typical sound level
65 dBA (measured 1 m from instrument)
Environmental ranges
Ambient temperature
15
℃
to 30
℃
Transport temperature
0
℃
to 45
℃
Storage temperature
15
℃
to 30
℃
Maximum relative humidity
80% (non-condensing)
Altitude (max.)
2000 m
Indoor use only
—
Not intended for use in a wet
location
—
Intended for use in Pollution
degree 2 environment
—
Operating limits
Centrifuge speed
0 to 3000 ×
g
(7000 rpm)
Flow rate
Standard: 5 to 110 mL/min (LS19)
Hi-Flow: 30 to 160 mL/min (LS16 tube)
System pressure
100 kPa
Liquid temperature
4
℃
to 38
℃
Fluid density
1.0 to 1.1 g/mL
Minimum working volume
Approx. 50 mL
Note:
Smaller volumes of input material can be
diluted to enable processing
Chapter 1
Product information
System overview
1
CTS
™
Rotea
™
Counterflow Centrifugation System User Guide
11
Содержание A44769