OPC-UA interface
Commands OPC-UA to Rotea
™
application
Table 2 OPC-UA command list
Variable
Type
Values/Description
RescanKit
Boolean
Force a kit rescan:
• True – The instrument will rescan the kit
barcode.
• False – No action.
SetGuiUserLevel
Enum
•
GUIManagedLogin
– Application login
is managed by the GUI, OPC-UA
commands are ignored in this mode.
•
ViewUser
– The MES has control of the
Rotea
™
instrument and GUI just displays
the instrument status.
•
AdminUser
– The GUI is logged in as
an administrator and has control the
Rotea
™
instrument, OPC-UA commands
are ignored in this mode.
Note:
When OPC-UA enabled, a
SetGuiUserLevel command from the OPC-UA
server will over-ride the current instrument log-
in state.
SetLocalProtocolFilePath
String
Set the file name path of the protocol on the
laptop to be loaded to the Rotea
™
instrument.
SetPlayPauseButtonEnable
Enum
Enable the Play/Pause button on the Rotea
™
instrument.
• Disabled
• AsPerProtocol
• Enabled
SetProtocolFileContents
String
Send the protocol file contents to the laptop to
be loaded to the Rotea
™
instrument.
SetProtocolFileByLookupTableId
UInt32
Id of the protocol file in the OPC-UA look
up table. The lookup table is defined in
OPCUAProtocolList.yaml
SetParameter1
through
SetParameter20
Int32
The Metadata parameter data for the protocol.
The associated name of the parameter can be
found in variables
ParameterName1
through
ParameterName20
after the protocol has been
loaded.
Chapter 4
OPC-UA interface
OPC-UA interface
4
96
CTS
™
Rotea
™
Counterflow Centrifugation System User Guide
Содержание A44769
Страница 131: ......