(continued)
Alarm or
message
Critical
alarm
Warning
Instrument response
GUI message
Over-pressure
condition
Yes
—
Trigger instrument stop state
Blockage detected
Protocol run has stopped.
Ensure manual clamps are
open and bags are connected
correctly. Press the stop button
to relieve pressure to Bag X.
Restart the protocol or select a
protocol step to start from.
Communication
error
—
Yes
Instrument continues running
protocol, GUI indicates error
until communications are
restored
Communication error
Ensure that the cable between
the laptop interface and
instrument is connected.
Kit removal
CAUTION!
Seal all fluid lines using a tube sealer or manual clamps prior to unlocking the door to
prevent fluid movement when the instrument pinch valves are released.
1.
Press
pushbutton when it is illuminated to unlock the door.
Chapter 3
Basic instrument operation
Kit removal
3
CTS
™
Rotea
™
Counterflow Centrifugation System User Guide
69