Table 2 OPC-UA command list
(continued)
Variable
Type
Values/Description
SetProtocolState
Enum
•
StopProtocol
– Stop the protocol.
•
SendProtocolToRotea
– Load the
protocol to the Rotea
™
instrument.
•
RunProtocol
– Start the protocol, same
as pressing the
Start
button on the
instrument.
•
PauseProtocol
– Pause the protocol
•
RecoveryRunProtocolFromStep
– Start
a protocol from the step specified in
SetRecoveryStepNumber
.
•
RecoveryRunProtocolFromStep
– Start
a protocol from the step specified in
SetRecoveryStepNumber
.
•
RecoveryRunProtocolSingleStep
– Start
a protocol and run a single step number
SetRecoveryStepNumber.
•
SkipToNextStep
- advances the protocol
to the next step by ending this step.
SetRecoveryStepNumber
UInt32
The recovery step number to start from. Used
by
SetProtocolState
.
SetSkipButtonEnable
Enum
Enable the advance button on the Rotea
™
instrument:
•
Disabled
•
AsPerProtocol
•
Enabled
Reports Rotea
™
application to OPC-UA
Table 3 Variable set when changed
Variable
Type
Value/Description
CentrifugeSpeed
UInt32
The current centrifuge speed
InstrumentName
String
The name of the instrument, stored on the
instrument, set by the GUI
MainFirmwareVersion
String
The main board firmware version on the
instrument
ModelNumber
String
The model number of the instrument, set
during manufacture
Chapter 4
OPC-UA interface
OPC-UA interface
4
CTS
™
Rotea
™
Counterflow Centrifugation System User Guide
97