background file name. Click the Use Specified Background File
checkbox and then enter the background spectrum filename.
If you have OMNIC Macros\Basic or OMNIC Macros\Pro, you can
create a macro that changes the angle incrementally with each
successive set of background and sample spectra.
Then use Experiment Setup to specify that the Macro button start the
experiment and run the macro. Use the Macro for Macro button field
on the Advanced tab in Experiment Setup dialog box (available
through Experiment Setup in the Collect menu) to specify the macro
to run. You may browse to locate the macro file or type the full path
and file name into the text field.
If you have OMNIC Series, you can display the series of spectra
collected in this way. To create a series (.SRS) file from your
spectra, follow the instructions in "Creating a series data set" in the
"Series Menu Commands" chapter of the OMNIC Series User's
Guide.
80 Thermo Electron Corporation