Problem
Possible Cause
Solution
The signal intensity is
low. (Also, the Scan
indicator may flash
intermittently and the
spectrometer alignment
may fail.)
The source is glowing
unevenly.
The beam path is obstructed.
The sample compartment
windows are fogged.
The beamsplitter is fogged.
The spectrometer is out of
alignment.
A sampling accessory is not
installed properly.
A detector that was just
installed is not seated properly.
Check to see if the source element is
glowing evenly. Order a
replacement source if needed.
Remove the obstruction.
Replace the windows. In
Spectrometer Help Topics view
“Replacing sample compartment
windows” in “Replacing parts” to
find part number, ordering and
installation instructions.
Check the laser signals by clicking
the laser indicator on the Diagnostic
tab of the Experiment Setup dialog
box.
For a Nicolet 6700
, replace the
beamsplitter, if necessary.
For a
Nicolet 4700
, contact technical
support.
Align the spectrometer using the
Align button on the Diagnostic tab
of the Experiment Setup dialog box.
Make sure the accessory is installed
according to the instructions that
came with it.
Make sure the hole and slot in the
detector plate fit over the pins on the
spectrometer baseplate. See
“Replacing the detector” in the
“Servicing your spectrometer” book
in Spectrometer Help Topics.
Nicolet FT-IR User’s Guide 143