Problem
Possible Cause
Solution
Laser status indicator is
off.
The beamsplitter compartment
cover is not fully closed.
Close the beamsplitter compartment
cover.
The laser is burned out.
Click the laser indicator on the
Diagnostic tab of the Experiment
Setup dialog box to verify that the
laser is burned out. Call Thermo
Electron to order a replacement
laser. View “Replacing the laser” in
the “Replacing parts” book in
Spectrometer Help Topics to find
part number and ordering and
installation instructions.
An error message
appears when you try to
scan.
The light source is not working.
An MCT detector is not cooled.
The interferogram peak is not
located at 1024 ± 16 data points
on the Bench tab of the
Experiment Setup dialog box.
The beam path is obstructed.
The data cable between the
spectrometer and the computer
is not properly connected.
The detector and beamsplitter
are not compatible.
Check the source indicator on the
Diagnostic tab of the Experiment
Setup dialog box. If necessary,
replace the source.
Check and cool the detector.
Align the spectrometer. If the
interferogram is still not located
correctly, call Thermo Electron for
service.
Remove the obstruction.
Turn off the spectrometer power and
check the data cable connections.
Change the detector or beamsplitter.
Nicolet FT-IR User’s Guide 147