You have just installed a laser, source, beamsplitter or detector.
•
The detector signal intensity has dropped significantly from its
usual level.
•
You have just moved the spectrometer.
?
In OMNIC Help Topics find “aligning, spectrometer” in the Index
and go to “Aligning the spectrometer.”
“Maintenance and Service” chapter...
“Aligning the spectrometer” in “Maintaining your spectrometer”
Does my detector need to be cooled with liquid
nitrogen?
Detectors that have a dewar need to be cooled. Dewar detectors are
fairly large metal cylinders with a hole in the top. These include
MCT and InSb detectors. To determine if your detector requires
cooling, open the detector compartment cover and see if the detector
has a dewar.
If during a work session you find that the detector signal intensity is
decreasing, the dewar may need to be refilled with liquid nitrogen.
Spectrometer Help Topics...
“Cooling a detector” in “Setting up experiments”
Nicolet FT-IR User’s Guide 125