Problem
Possible Cause
Solution
The baseline is not
stable.
The purge flow rate is too high.
The spectrometer is out of
alignment.
The laser detector is
misaligned.
The ambient conditions are
outside the allowed ranges.
The desiccant cannot absorb
any more water vapor.
The spectrometer cover was
recently opened.
Lower the flow rate until the
baseline is stable.
Align the spectrometer using the
Align button on the Diagnostic tab
of the Experiment Setup dialog box.
Use the Bench Diagnostics software
to check the laser levels for proper
balance. If the levels are out of
tolerance, align the spectrometer
using the Align button on the
Diagnostic tab of the Experiment
Setup dialog box. If the levels
remain out of tolerance, call Thermo
Electron for service.
Make sure the conditions meet the
requirements given in the
“Environmental Considerations”
chapter of
Preparing Your Site
.
Check the humidity indicator and
replace the desiccant and indicator if
needed.
Allow the spectrometer to purge for
3 to 5 minutes after the cover is
closed.
Nicolet FT-IR User’s Guide 139