Warning
Do not operate the spectrometer with the main cover removed. The
cover protects you from exposure to laser light and live electrical
connections.
S
If you must open the main cover for maintenance or service reasons,
always turn off the spectrometer power and unplug the power cord
first.
Important
If you have a CsI beamsplitter, protect it from the effects of water
vapor by minimizing the time that the cover is removed.
S
Spectrometer Help Topics...
“Opening the main cover” in “Installing optional hardware”
When should I align the spectrometer?
Align the spectrometer in the following situations:
•
The ambient temperature has changed by at least 5 Celsius
degrees (about 9 Fahrenheit degrees).
•
You are about to perform a spectrometer validation with Val-Q
(or other software). This ensures that the validation is run with
the best possible interferometer and electronics settings.
•
The spectrometer power was turned off for more than a few
minutes and has now been on for one hour.
Note
We recommend that you leave the power on continuously, but if you
do need to turn it off, allow the spectrometer to warm up for one
hour and then align it before collecting spectra.
S
•
The spectrometer has been on continuously for one week without
being aligned. We recommended aligning your spectrometer
once a week or, for optimum performance, once a day.
124 Thermo Electron Corporation