When should I change the desiccant (in sealed and
desiccated spectrometers only)?
If your spectrometer is in use
, check the performance of the
desiccant by using the Check Desiccant button on the Diagnostic tab
of the Experiment Setup dialog box. If the check indicates that the
desiccant is no longer effective, replace it.
Spectrometer Help Topics...
“Checking and changing desiccant” in “Maintaining your
spectrometer”
If your spectrometer is in storage
, check the round, blue indicator on
the desiccant holder under the beamsplitter compartment cover
monthly. Change the desiccant if the indicator has turned pink.
See “Ordering parts” in Spectrometer Help Topics to find the part
number for ordering desiccant bags. The desiccant holder under the
beamsplitter compartment cover holds one bag.
Which source should I use?
Use the source that is appropriate for the spectral range of your
experiment (determined by the ranges of the beamsplitter and
detector). Thermo Electron offers a mid-IR to far-IR source (ETC)
and a near-IR source (white light).
?
In OMNIC Help Topics find “source” in the Index and go to
“Specifying the source type.”
“Spectrometer Basics” chapter...
“Selecting the right beamsplitter and detector” in “Using
beamsplitters and detectors”