Using the
internal polarizer
An optional automated polarizer lets you automatically control the
orientation of the electric field vector of the IR beam entering the
sample compartment. The range of orientation is 0 to 180 degrees
in increments of 1 degree. At 0 degrees, the electric field vector is
oriented vertically, perpendicular to the sample compartment base
plate. This option is useful for performing polarization studies of
thin films and other sample types.
Note
The polarizer cannot be installed on spectrometers with sample
compartment windows.
S
In Spectrometer Help Topics view “Installing optional hardware” for
detailed information about installing the polarizer.
In these experiments, a unique background is required for each
polarization position used. For example, if you are going to collect
spectra at 0 and 90 degrees, corresponding background spectra are
required at 0 and 90 degrees.