Fields:
<CalDate> (TimeDate) is the date and time the calibration was performed.
<NextCalDate> is the date and time the next calibration is due.
Returns:
+OK:~: - The dates and times were set successfully
+ERROR:0006:~: - Calibration Jumper not inserted
+ERROR:0007:~: - Error writing the data
+ERROR:0009:~: - One of the parameters was invalid
Comments:
The system checks the validity of both TimeDate Structure:s before writing them to
the memory. If either are invalid, the data is not written and the function returns an error
Setup:calreadInfo
This reads the calibration information that is currently stored in the instrument.
Structure:
+S:I:~:
Fields:
None
Returns:
+OK:<Calibrator>:<CalDate>:<NextCalDate:~:
<Calibrator> is the identity of the last person to calibrate the system
<CalDate> (TimeDate) is the date of the last calibration held in EEPROM
<NextCalDate> is the date of next calibration held in Battery powered memory.
Comments:
This allows the user to query the calibration information before and after it is set up.
The values read back depend on whether they have been updated by the user during the calibration session. If the
values are changed, then those changes will be reflected in the data returned when the calreadInfo function is called.
Setup:CurrentTimE
This sets the real time clock
Structure:
+S:E:<TimeDate>:~:
Fields:
<TimeDate> is the time and date to be set
Returns:
+OK:~: - The time was updated successfully
+ERROR:0009:~: - The TimeDate Structure: was invalid.
Comments:
None
Setup:SetGain
This sets the gain of the measurement system
Structure:
+S:G:<Gain>:~:
Fields:
<Gain> (integer) is the bit wise combination of the following constants as required:
MEAS_FILT60
(0x80)
- Sets 60Hz filter
MEAS_PRIAUTO
(0x70)
- automatic gain determination
MEAS_PRIX1 (0x00) - x1 Primary Gain
MEAS_PRIX8 (0x30) - x8 Primary Gain
MEAS_SECAUTO
(0x0F)
- automatic gain determination
MEAS_SECX1
(0x00)
- x1 Secondary Gain
MEAS_SECX2
(0x01)
- x2 Secondary Gain
MEAS_SECX4
(0x02)
- x4 Secondary Gain
MEAS_SECX8
(0x03)
- x8 Secondary Gain
MEAS_SECX16
(0x04)
- x16 Secondary Gain
MEAS_SECX32
(0x05)
- x32 Secondary Gain
MEAS_SECX64
(0x06)
- x64 Secondary Gain
MEAS_SECX128
(0x07)
- x128 Secondary Gain
Содержание TTR 2795
Страница 1: ...Operating Instructions HAEFELY TEST AG TTR 2795 Transformer Turns Ratio Meter Version 1 4 ...
Страница 10: ...6 Introduction Maximum Roll Diameter Outer 50mm Inner 10mm ...
Страница 22: ...18 Connection and Setup Figure 11 Connecting to three phase BCT ...
Страница 101: ...Application Software APSW 2795 97 ...
Страница 111: ...Miscellaneous 107 14 7 Conformity ...