Remote Control
61
determination. The valid voltage levels for the test are:
0x0A (10) - 10 Volts energization
0x28 (40) - 40 Volts energization
0x64 (100) - 100 Volts excitations
Returns:
+OK:<VectorGroup>:<TestVoltage>:~: - Vector group updated
+ERROR:0909:~:
- Vector group is invalid
+ERROR:090A:~: - Voltage is invalid
+ERROR:0902:~:
- Working Memory has data
+ERROR:0300:~:
- No update - test running
The vector group is returned in the response as a confirmation. If there was a
problem
with the vector group, the function gives a negative response and returns the actual
vector
group being used. The function also returns the test voltage to be used. If an invalid
test
voltage is specified, a positive result will still be generated (as long as the vector
group is
valid), and the test voltage set to automatic determination.
Comments:
The vector group will be updated as the test runs, the vector group reported at the
end of the testy may differ from that at the start of the test, especially if automatic determination is being used.
Test:Setup:Taps
This sets up the number of taps, the bottom tap, the nominal tap, the tap setup and the step voltage value on the
system. The HV Nominal Voltage and LV Nominal Voltage have to be set before setting the Tap values.
Structure:
+T:S:T:<NumTaps>:<BotTap>:<NomTap>:<StepValue>:~:
Fields:
<NumTaps> (integer) is the number of taps on the transformer. A value of zero
indicates
no taps on the transformer. A value of one indicates a single tap (i.e 2 outputs) and
so on.
<BotTap> (integer) is the number of the bottom tap on the transformer.
<NomTap> (integer) is the index of the Nominal Tap on the transformer.
<StepValue> (float) is the tap-to-tap Step value. This is in percentage or voltage
deviation
from the Nominal Voltage. If the tap is on the HV side of the transformer, Step
value is
negative. If the tap is on the LV side of the transformer, step value is positive. For
manual
setup, the step value is zero. Based on the values of the the stepvoltage or
percent, the
tapsetup field is updated as HV or LV or Manual.
Returns
:
+OK:<NumTaps>:<BotTap>:<> NomTap:< StepValue >:~: -Taps updated
+ERROR:0902:~:
- Working Memory has data
+ERROR:0907:~:
- Number of taps out of range
+ERROR:090B:~:
- Bottom tap out of range
+ERROR:0915:~:
- Invalid Step Percent value
+ERROR:0916:~:
- Invalid Step Voltage value
+ERROR:0917:~:
- Nominal Tap out of range
Содержание TTR 2795
Страница 1: ...Operating Instructions HAEFELY TEST AG TTR 2795 Transformer Turns Ratio Meter Version 1 4 ...
Страница 10: ...6 Introduction Maximum Roll Diameter Outer 50mm Inner 10mm ...
Страница 22: ...18 Connection and Setup Figure 11 Connecting to three phase BCT ...
Страница 101: ...Application Software APSW 2795 97 ...
Страница 111: ...Miscellaneous 107 14 7 Conformity ...