Remote Control
73
0x0001
System has printer. If this flag is set the system has a printer available,
so should print test reports.
Setup:SystemType
This sets the system type for the unit
Structure:
+S:T:<Type>:~:
Fields:
<Type> (string) is the type of the instrument. This is typically TETTEX2795 or
TETTEX2795R
Returns
:
+OK:~: - The System type was set successfully
+ERROR:0006:~: - Calibration Jumper not inserted
+ERROR:0007:~: - Error writing the data
Comments
:
This parameter is used on the test report and as part of the startup screen, to
indicate the instrument type. If the string is longer than 16 bytes including the terminating null, it is truncated and the
terminating null placed in the last place.
Setup:SerialNo
This sets the serial number of the unit
Structure:
+S:S:<Serial No>:~:
Fields:
<Serial No> (string) is the serial number of the unit.
Returns:
+OK:~: - The System type was set successfully
+ERROR:0006:~: - Calibration Jumper not inserted
+ERROR:0007:~: - Error writing the data
Comments:
This parameter is used on the test report and as part of the startup screen, to
indicate the identification of the instrument. If the string is longer than 16 bytes including the terminating null, it is
truncated and the terminating null placed in the last place.
Setup:calibrator
This sets the identity of the last person to calibrate the system
Structure:
+S:A:<Calibrator>:~:
Fields:
<Calibrator> (string) is the serial number of the unit.
Returns:
+OK:~: - The System type was set successfully
+ERROR:0006:~: - Calibration Jumper not inserted
+ERROR:0007:~: - Error writing the data
Comments:
This parameter is used to identify the technician/engineer who last calibrated the
instrument. If the string is longer than 16 bytes including the terminating null, it is truncated and the terminating null
placed in the last place.
Setup:calDateTime
This sets the date and time of this calibration and the due date of the next one.
Structure:
+S:D:<CalDate>:<NextCalDate>:~:
Содержание TTR 2795
Страница 1: ...Operating Instructions HAEFELY TEST AG TTR 2795 Transformer Turns Ratio Meter Version 1 4 ...
Страница 10: ...6 Introduction Maximum Roll Diameter Outer 50mm Inner 10mm ...
Страница 22: ...18 Connection and Setup Figure 11 Connecting to three phase BCT ...
Страница 101: ...Application Software APSW 2795 97 ...
Страница 111: ...Miscellaneous 107 14 7 Conformity ...