28
Operation Procedure
Turning the Power On
Please ensure the product is fit for purpose prior to use.
Cautions
• Conduct the self-check (self-diagnosis) of this product without a tube installed and with the door left open. Otherwise the
self-check may not be conducted properly.
• Check for any fault by conducting a self-check. If a fault is observed, immediately discontinue use and contact TERUMO
trained service technicians.
1 second
or more
•
Keeping the door open, press and hold the Power switch
for 1 second or more until the TERUMO logo screen is
displayed to turn the power on.
Checks
When turning the power on, the following occur simultaneously;
• Fingers move slightly
• TERUMO logo appears on the display
• LCD flashes 3 times
• Operation indicator repeats flashing red and green alternately, and then buzzer sounds.
Operation indicator
Fingers
Notes
• A timer for maintenance can be set in this product. When the period set for the
maintenance timer is reached, a maintenance request is displayed at start up.
When the request is displayed, contact TERUMO trained service technicians.
• When the period set for the maintenance timer is reached, this screen displays
at every start up until the setting is changed. For the setting method, contact
TERUMO trained service technicians.