43
Operation Procedure
Close
3)
Close the manual roller clamp on the infusion set.
Door lock lever
Door
Release lever
Tube clamp
4)
Open the door, press down the release lever to release the
tube clamp, and remove the tube from the tube guides and
tube holder.
Warning
• Ensure the manual roller clamp is closed on the infusion set before opening
the door, and then release the tube clamp to remove the tube.
Turning the Power Off
Power switch
3 seconds
or more
•
When the operation ends, press and hold the Power switch
for 3 seconds or more to turn the power off.
Notes
• While holding the Power switch, the power off buzzer sounds, then the LCD
displays count-down arrows and turns off.
• If the Power switch is pressed and held during solution delivery, a buzzer sounds
indicating that solution delivery is in progress. The buzzer continues to sound for
approx. 3 seconds until the power is turned off.