85
In order to ensure the safe use of the infusion pump, make sure it is cleaned and inspected after use and stored in an appropriate
environment. If you find any damage or malfunction, immediately stop using the pump and contact TERUMO trained service
technicians.
Cleaning
Cautions
• Clean the AC inlet, fingers, occlusion detector etc. on a regular basis. Failure to conduct the regular cleaning may result in
damage, failure or malfunction to/of this product.
• If any parts (fingers, occlusion detector etc.) of this product need to be removed for cleaning, contact TERUMO trained
service technicians.
• When disinfecting, do not use a sterilizer. Use a gauze (damp with antiseptic solution) to wipe the product, and wipe off the
antiseptic solution with a gauze (damp with cold/lukewarm water), and then thoroughly wipe off any moisture with a dry soft
cloth.
• Before cleaning, always turn off the power and disconnect the AC power cable. Otherwise it may cause an electrical shock
hazard. When AC power is supplied by the optional rack system (Model: TE-RS700), first remove the pump from the rack
system, then conduct cleaning.
• If any adhered drug solution is present, the solution delivery or alarm detection may not operate correctly or may interfere
with the normal operation of the clamp part and/or movable part of the pole clamp. If any drug solution is present, clean
immediately with a cotton swab.
• Do not wipe with undesignated alcohol, other organic solvents such as thinner, or povidone-iodine. Using organic solvent
or any antiseptic solution other than those permitted for use may result in damage to or failure of this product.
• Do not use an electric dryer to dry this product. This product may be damaged.
• This product does not have a waterproof structure. Therefore, do not wash with running water or immerse in water. It may
cause damage or failure of the product.
Examples of disinfectants (ingredient names) which can be used are listed below.
Examples of Cleaning Disinfectants (Ingredient Names)
Ingredient Name
Dilution (e.g.)
Chlorhexidine gluconate
5%
Benzalkonium chloride
10%
Ethanol*
76.9 - 81.4vol%
When using disinfectant, follow the instruction manual of each disinfectant (regarding degree of dilution, etc.).
* Do not use ethanol to disinfect the following:
• Air-in-line detector and occlusion detector of the main unit
• Drip sensor
• One Touch Pole Clamp (specified parts only)
For details, see page 86.
After Use