30
Operation Procedure
Close
4)
After the completion of priming, close the manual roller
clamp.
Attaching a Tube
1)
Check
• The infusion set to be used matches the setting of the main unit.
2)
Press down the release lever to open the tube clamp.
Tube guides
Fingers
Lower
occlusion
detector
Upper occlusion
detector
Air-in-line detector
Upper
side
Lower side
3)
Keep the tube straight and insert the tube completely into
the tube guides (4 positions).
Checks
• The tube is not broken or crushed, and that it is set straight without sag.
• The tube is correctly set in the air-in-line detector and the occlusion detectors.
Example of false installation
Tube clamp
Lower occlusion
detector
Upper occlusion
detector
Air-in-line detector
Cautions
• When setting the tube, ensure that there is no bend, dent or slack in the tube and that the tube is set straight into the
fingers, tube guides, various detectors and tube clamp. Incorrect loading may result in abnormal solution delivery such as
excessive infusion, insufficient infusion or non-administration of drug solution.
• When setting the tube, do not pull excessively hard. The tube may be deformed, and therefore the original functionality or
performance of this product may not be achieved. (flow rate error and various alarm functions etc.)
• Do not use the tube in the same condition for more than 24 hours. It is recommended either to slide the tubing by at least
15 cm once in every 24 hours or replace with a new infusion set.
If the tube is used for a long period, it may be deformed resulting in a flow rate error or Occlusion alarm.