25
Preparation
Connecting to AC Power
AC power cable
1)
Firmly connect the AC power cable (included) to the AC inlet.
2)
Connect the plug an AC power source with an earth
connection.
Note
• If an AC power cable is not included with the pump unit, please contact TERUMO
trained service technicians to receive the cable suited for your region.
Check
• The AC icon and battery icon should be displayed on the LCD.
Notes
• Even if the power is off, if the AC power has been supplied, the AC icon is displayed.
• The battery can be charged even if the power is off. The battery icon
(charging) is displayed. There may be cases where the
remaining battery level and
(battery failure) flash alternately every other second when there is a sub-battery failure or low
battery.
• If the AC power cable is disconnected or connected with power turned on, the buzzer sounds.
• If the AC power cable is disconnected with power turned on, the LCD becomes dim (2 levels of brightness setting). When the AC
power cable is reconnected, the previous brightness is restored.
Cautions
• Please ensure AC inlet and all connections are clean and dry prior to use. If moisture is present, make sure that the power
is turned off and the AC power cable is removed from both main unit and AC power source, and then thoroughly wipe with
a dry cloth.
• Use the AC power cable included and connect to an earthed AC power source.
• Use an earthed AC power supply for normal use. The internal battery is an auxiliary power source for when the AC power
supply is not adequately available during transportation, power failure, etc.
• This product should only be used with a sufficient power supply. If the power supply is not sufficient, the internal battery
will be used for the operation, which may result in no power supply in an emergency situation.