101
Troubleshooting
Start Reminder
Occurrence
Cause
Action
• LCD: The Start Reminder
display lights up yellow..
• Operation indicator: Off
• Buzzer: Activates.
• Pump: Remains stopped.
• The product has been left for
2 minutes or more after being
programmed.
• If you want to delay operation for a while, press and
hold the Back/Mute switch for 2 seconds or more to set
the equipment to standby mode. (See page 47.)
• Check the setting values for flow rate and VTBI,
and press the Start switch to start solution delivery.
(See pages 35 and 36.)
Battery alarm
Occurrence
Cause
Action
• LCD: The Battery alarm
display lights up yellow.
• Operation indicator: Flashes
green (during solution delivery)
or off (stopped).
• Buzzer: Activates.
• Pump: Continues solution
delivery (during solution
delivery) or remains stopped
(stopped).
• Low battery capacity
(approximately 30 minutes),
battery level low or battery
energy is reducing.
• Connect to the AC power supply or TERUFUSION
Standard Rack System. (See page 25.)
* When using a rack system, see the instruction manual
of the rack system.