3.
Unit capacity too low. This occurs when the process
requires more heat than the unit is capable of producing.
The only option in such cases is to install a unit with an
adequate heater KW rating for the load.
4.
Instrument defective. The instrument is designed and
manufactured exclusively by Temptek. The instrument is
life-tested and found to be field reliable. However, in the
case where the instrument is determined to be defective,
the operator must remove the assembly according to
instructions outlined in this manual and return the assembly
to the factory. The instrument is not a field serviceable
component.
4.6
PRESSURE RELIEF VALVE LEAKS
A.
The unit has a 150 psi pressure relief valve mounted in the cooling
cylinder. If the valve is found to be leaking, the operator should
check the following:
1.
Water supply pressure exceeds 75 psi. The unit is
designed to operate with water supply NOT exceeding
75 psi. See section 2.3 paragraph B for specific water
supply pressure requirements at corresponding setpoint
temperatures. If the plant water supply pressure exceeds 75
psi, the pressure relief valve may leak. Static water supply
pressure can be determined at the unit’s location by reading
the unit’s 0-160 PSI pressure gauges when the unit’s
motor pump is off. If the water supply pressure at the unit’s
location exceeds 75 PSI, a pressure reducing valve must be
installed in the water supply line. The factory recommended
‘regulated pressure out’ is 55 PSI. Refer to section 7.4 for
regulator installation drawing.
2.
Back flow prevention device in water supply line. If a back flow
prevention device (check valve, pressure regulator, closed valve) is
installed in the water supply line, increased pressures from thermal
expansion are unable to move into the water supply line. This will
increase the unit’s internal pressure causing the pressure relief
valve to leak. Refer to section 7.4 for regulator installation drawing.
3.
Valve contamination. The pressure relief valve may become
contaminated with water debris causing the valve not to close
properly. If this is the case, flushing the valve for a moment will
cleanse the seat and allow it to work properly. If flushing the valve
does not remedy the leaking, the valve must be replaced.
4.
Extreme internal system pressure. If the internal pressure in
the Sentra unit is elevated, the pressure relief valve will leak as a
safety measure to dissipate excessive pressure. If this is the case,
the operator must determine why the system internal pressure is
excessive and correct the condition.
Temperature Control Units : With ‘LS’ Series Instrument
Page: 33
TEMPTEK, INC.
525 East Stop 18 Road Greenwood, Indiana 46142
317-887-6352 Fax: 317-881-1277
Service Department Fax: 317-885-8683
www.Temptek.com
Содержание VETERAN LS Series
Страница 2: ......
Страница 8: ...Page 8 THIS PAGE INTENTIONALLY BLANK ...
Страница 60: ...Page 60 THIS PAGE INTENTIONALLY BLANK ...
Страница 68: ...Page 68 THIS PAGE INTENTIONALLY BLANK ...
Страница 78: ...Page 78 THIS PAGE INTENTIONALLY BLANK ...
Страница 79: ...END 2013 Temptek INC RE 1 02 13 ...
Страница 80: ......