temperature on the display should equal the temperature of
the ice water bath as indicated by the digital thermometer. If
not, the operator must change the calibration of the probe.
a.
To access the calibration potentiometer, open
the electrical cabinet panel door. The panel door
is secured by a support strap. Caution must be
employed when the electrical panel door is open
since power is applied to the unit.
b.
Locate the instrument
CPU. The calibration
pot is located on the
“mother board” of the
instrument assembly
(figure 5.4C).
c.
Use a non-conductive
device, to adjust
the potentiometer.
Adjust the
potentiometer until
the “to process”
temperature on the display equals the temperature
of the ice bath.
9.
When the two temperatures (“to process” and ice water
bath) are equal, the calibration procedure is complete.
10.
Shut down the unit. The operator must be certain to
remove the 1/2” plug and replace the sensor probe. Restart
operations.
Calibration port
Figure 5.4C