EuroFlow EF and EF/B
Biological Safety Cabinet
UM-EFAB-EN10.doc
53
Figure 27 Fold the worktop against the back partition
Clean the sides and bottom part of the worktop;
Fold the worktop back in place.
Cleaning the inside of the window
Check if this procedure can be done safely. It is advised to disinfect the installation first (see
Appendix II: Disinfection on page 67
).
Procedure for cleaning and to disinfect the window with liquid:
Put on gloves;
Switch off the installation (see
§5.7 Turn off on page 28
);
Hinge the window open (see
§5.2 Hinge the window on page 22
);
Clean (and, if necessary, disinfect) the window;
Clean (and, if necessary, disinfect) the rubber seal;
Smear the seal thin with Vaseline;
Hinge the window close.
Please note: the cleanability of the biological safety cabinet is at least conform Class CI-B (EN
12469:2000-table 2: Cleanability performance).
7.2
Cleaning the UV-light (optional)
BIOLOGICAL DANGER
Never clean an used installation without precaution
(for instance disinfection)
DANGER FOR ELECTROCUTION