EuroFlow EF and EF/B
Biological Safety Cabinet
UM-EFAB-EN10.doc
12
In case of maintenance the cabinet must be completely free of voltage. This means that the
connector in the socket must be easy to reach, otherwise a main-switch must be mounted
between the cabinet and the power supply so that the cabinet can be cut off of the power-circuit.
3.2
Transport instructions
Before moving the cabinet make sure the power cable is disconnected of the power supply
The cabinet must be transported on a strong solid pallet.
After every transport the cabinet must be tested and calibrated.
GENERAL DANGER
Be careful: High center of gravity, crossover point.
3.3
Dismantling & Disposal
For dismantling and disposal follow the next instructions:
•
See
§7.1 Cleaning the installation on page 51;
•
See
§8.2 Replacement of the pre-filter on page 56
;
•
See
Appendix II: Disinfection on page 67;
•
See
Appendix III: Replacement of the HEPA filters on page 68;
•
Break the unit down into its component parts. You can re-cycle these components parts,
dispose them in accordance with local requirements.
For advice or information about dismantling or disposal please contact Clean Air Techniek B.V.