EuroFlow EF and EF/B
Biological Safety Cabinet
UM-EFAB-EN10.doc
38
There are three ways to set the
sockets:
-
Timer
-
7 days clock
-
Disable
2
See
§5.11 Set up 7 days clock
UV on page 33 and further.
See also the
Example for
programming the UV-lamp on
page 34.
The timer function needs to be activated first, otherwise the
sockets will not switch on.
Set up cabinet 7 days clock
With this menu it is possible to program an automatic switch action on of the cabinet. It is
however not possible to switch off the cabinet itself automatically!
1
Cabinet in work mode.
See
§5.4 Starting up the
cabinet on page 26.
•
Push 5x button
(1x Menu, 1 x A and 3x =>),
•
Acknowledge the setting with button
.
Cabinet in work mode - Safe
Monday 11:06:33
MENU
Socket=Off
Setting cabinet 7 days clock
Function: disabled
A = Enter
=>
<=
Esc
Tuesday
Cabinet to work mode at 0.00
Disable
Change
Esc
2
Disable
•
Push button
(Disable), to reset all previous
settings.
3
Change days:
•
Push button
(Change), to change the day.
•
The value can be changed by pushing,
(+) or
(-).
Tuesday Change day
Cabinet to work mode at 0.00
A = Enter
>>
+
-
Esc
4
Change hours:
•
Push button
(>>) to change the time.