EuroFlow EF and EF/B
Biological Safety Cabinet
UM-EFAB-EN10.doc
20
•
The connection, that is located outside the installation, is standard assembled in the bottom of
the installation.
•
The connection, that is located outside the installation, can be assembled by customer request
in left or right side panel or top of the installation.
Positions of the supply connection EF/B
•
The connection, that is located inside the installation, is standard assembled in combination
with a vacuum tap in the back panel of the workspace.
•
The connection, that is located outside the installation, is standard assembled in the left or
right side panel of the installation.
•
The connection, that is located outside the installation, can be assembled by customer request
in top of the installation. For EF/B-series assembly of this connection on the bottom is not
possible.
UV light
The UV light is meant for disinfection of the working surface of the safety cabinet.
When the UV light switches on, the FL lighting switches off. The UV light can only be switched on
when the window is closed (window on aerofoil) or when the cabinet is in disinfection mode
(aerofoil lifted up and window at its lowest point), this makes sure the user will not get in contact
with UV-C radiation.
The UV switch-function
is only visible when the window is closed.
DANGER FOR BURNING
Avoid direct contact with UV-C radiation. This causes
burns to the skin and eyes.
The UV light can be switched on and off on the display (see
§5.8 Control UV (optional) on page
29
). It is also possible to adjust the UV disinfection, for each day a certain disinfection period can
be defined (see
§5.11 Set up 7 days clock UV on page 31)
.
Watch Out!
Do not touch the UV-lamp with the hand and keep the
lamp fat free; otherwise the lifetime will decrease. The
UV-light must be cleaned with alcohol (70%) regularly
(switch off the installation). See
§7.2 Cleaning the UV-
light (optional) on page 53.