Screen II Installed options – relays
Installed Options
Relais #A1
Relais #B1
=>
<=
Esc
Figure 47 Example of screen II
Overview functions of relay K8, #A1-A12
Alarm #A1 - #A12 relay switching K8, on output J5 at the IO print:
Set up from #A1 until #A12:
A1
Relay ON = A & ( HD * LD * HI * LI )
DEFAULT
A2
Relay ON = A & ( HD * LD * HI * LI * R )
A3
Relay ON = W & ( HD * LD * HI * LI )
A4
Relay ON = W & ( HD * LD * HI * LI * R )
A5
Relay ON = W & ( HD * LD )
A6
Relay ON = A & ( HD * LD )
A7
Relay OFF = A & ( HD * LD * HI * LI )
A8
Relay OFF = A & ( HD * LD * HI * LI * R )
A9
Relay OFF = W & ( HD * LD * HI * LI )
A10
Relay OFF = W & ( HD * LD * HI * LI * R )
A11
Relay OFF = W & ( HD * LD )
A12
Relay OFF = A & ( HD * LD )
Explanation codes used above:
W = Working mode, S = Save mode, A = Cabinet on and spare mode, HD = High downflow alarm,
LD = Low downflow alarm, HI = High inflow alarm, LI = Low inflow alarm, R = Window alarm, & = AND function,
* = OR function.
Functions are factory adjustments and cannot be changed by the user.
Overview functions of relay K3, #B1-B5
Alarm #B1 - #B5 relay switching K3, on output J22 at the IO print:
•
#B1: the K3 relay is activated if the main fan is switched on (working mode / saving mode);
•
#B2: the K3 relay is activated if the main fan is on half speed mode (saving mode);
•
#B3: the K3 relay is activated if the main fan is in the working mode;
•
#B4: the K3 relay is activated if the option Disinfection in the menu is chosen;
•
#B5: the K3 relay is activated if there is an alarm from the analogue inputs.
Overview functions of relay K2, #C1-C5
Alarm #C1 - #C5 relay switching K2, on output J14 at the Option print:
•
#C1: the K2 relay is activated if the main fan is switched on (working mode/ saving mode);
•
#C2: the K2 relay is activated if the main fan is on half speed mode (saving mode);
•
#C3: the K2 relay is activated if the main fan is in the working mode;
•
#C4: the K2 relay is activated if in the menu is chosen for the option Disinfection;
•
#C5: the K2 relay is activated if there is an alarm from the analogue inputs.