EuroFlow EF and EF/B
Biological Safety Cabinet
UM-EFAB-EN10.doc
43
•
Preparations for disinfection
•
Activating the user code
•
Factory adjustments
•
Testing of alarms
•
Change the user code
•
Change language
Disinfection
Your Biological Safety Cabinet has a “built-in” disinfection software program.
This program knows a number of varieties, which are dependent on the installed options in the
security cabinet, and the manner the cabinet has been installed.
The disinfection software of the control print is only available in version number 2.4 270307 or
higher.
The disinfection software itself is no disinfection!
The
Biological Safety Cabinet EF/S can work according to automatic disinfection software only if
this has been installed.
Consult your order affirmative for the installed options or contact Clean Air B.V.
In case additional information about Disinfection is needed, please contact Clean Air B.V.
Also check the site of Clean Air B.V. (
www.cleanair.eu/products/downloads
) for more detailed
information about disinfection.
Depending on the code of the installation the safety cabinet will react differently. These codes are
programmed for an optimised disinfection process. The code itself is already pre-installed on the
cabinet depending on the installed options mentioned in your order affirmative or contact Clean Air
Techniek.
•
Check if the cabinet is gastight.
•
Use the valid standard operation procedure (SOP) since the disinfection procedure depends
on the handlings performed in the cabinet.
1
Cabinet in work mode.
See
§5.4 Starting up the cabinet
on page 26.
•
Switch off the cabinet with button
.
•
Push button
(Menu).
•
Push button
(C).
•
Push button
.
Cabinet in work mode - Safe
Monday 11:06:33
MENU
Socket=Off
A => Hour counter information
B => Status cabinet
C => Settings
A
B
C
Esc
2
Preparations for disinfection:
Preparation for disinfection display shows.
•
After entering the user code (pre settings 0001) push
Preparations for disinfection
0000 (Enter code)
A = Enter
=>
+
-
Esc