Configure the Instrument
Set and Recall Instrument Presets
PRISM MPI2-10-MPX2-10 User Manual
128
Gamut
YRGB Gamut: Indicates when the input signal contains colors that violate the currently set limits for
YRGB gamut, which are specified by the Settings/Alarms/Gamut. These measurements conform to
the EBU R103 standard.
Gamut measurements are supported for SD, 720p,1080p slow frame rate, 1080i, and 1080p
levelA 422-10bit and 444-12bit.
Gamut measurements can be totally disabled to minimize CPU loading to enable quieter
operation. To disable the Gamut measurements, set Gamut Enable to
Off
.
Note:
When Gamut is disabled, no Gamut alarms are generated or logged.
Note:
The Gamut Preset button, EBU R103, only resets the YRGB numbers, not the picture area or
frame duration.
IP
•
Combined RTP Sequence Error: RTP sequence error occurred in the combined output of a media
stream protected using RTP packet reordering or S2022.7 seamless switching.
•
Data Rate Invalid or Unstable: Detected data rate of a video input is unstable or nonstandard.
•
Ethernet BER Excessive: Excessive bit-error rate has been detected on Ethernet interface.
•
Ethernet Buffer Overflow: Overflow was detected on IP Video Ports.
•
Ethernet CRC Error: CRC error was detected in an Ethernet frame.
•
Ethernet Interface Saturated: Ethernet interface is saturated.
•
HBRMT Header Invalid: Error occurred in the HBRMT header.
•
HBRMT Sequence Error: Packets are dropped, out of order, or duplicated; or a sequence number in
a flow does not change.
•
HBRMT Timestamp Error: HBRMT timestamp is corrupt or is not updating at the correct frequency
for the specified media type.
•
L4 Protocol Checksum Error: Checksum error was detected in an IP or UDP packet.
•
Network Connection Lost: Fiber connection to SFP port has been removed or FEC mode is in the
wrong state.
•
RTP Header Invalid: Error occurred in the RTP header.
•
RTP Marker Frequency Error: RTP marker bit is not designated at the correct frequency for the
specified media type.
•
RTP Sequence Error: Sequence errors can be caused by dropped, out of order, or duplicated
packets. They can also be triggered when the sequence number within a flow does not change.
•
RTP Timestamp Error: RTP timestamp is corrupt or is not updating at the correct frequency for the
specified media type.
•
S2022/7 Path Missing: One of the datapaths in a S2022.7 seamless switching configuration is
missing.
Group alarm titles (bold), related alarms, and the related-alarms descriptions
Содержание PRISM MPI2-10
Страница 1: ...PRISM MPI2 10 and MPX2 10 SDI IP Waveform Monitor User Manual April 2021 D00010021C...
Страница 6: ...6 PRISM MPI2 10 MPX2 10 User Manual...
Страница 52: ...Getting Started Activity Center PRISM MPI2 10 MPX2 10 User Manual 54...
Страница 74: ...Methods of Operation Remote Control through API Commands PRISM MPI2 10 MPX2 10 User Manual 76...
Страница 302: ...PTP Operational Overview Compensating for Causes of Asymmetric Delay PRISM MPI2 10 MPX2 10 User Manual 304...