Operator’s Manual Simple AFG Application
70
Clock Settings tab
The Timing settings parameters are located in the Clock Settings tabs.
Timing Setting
Description
Clock Source
If
Internal Clock
is selected, the internal DAC sampling clock is synthesized
using a 10Mhz reference clock generated internally.
If the external
Reference Clock In
(
REF CLK
) is selected, the DAC sampling
clock is synthesized using the clock provided externally to the Ref.Clock In
SMA connector on the rear pannel of the instrument. In this case the
Ref Clk
Frequency
control will appear and the user must specify the Reference Clock
frequency [Hz]
.
The frequency range of this clock is: 5 Mhz to 100Mhz.