Operator’s Manual Simple AFG Application
72
Edge
option is the opposite.
“
Both Edges
” means that Trigger is sensitive to both edges of the signals.
It has effect only when the selecting Source is
Trigger In Connector
.
Input Impedance
It selects the Trigger In connector impedance: 1 kOhm or terminated into 50
Ohm.
Marker Settings
In the marker output page, it is possible to define the behaviour and parameters of the Marker Out
signals located in the front panel of the instruments.
The T3AWG-3352/3252 model has one Marker Out: Marker Out 1. The T3AWG-3354/3254 model has
two Marker Out: Marker Out 1, Marker Out 2. The T3AWG-3358/3258 model has four Marker Outs:
Marker Out 1, Marker Out 2, Marker Out 3, Marker Out 4.
Each Marker Out has its own set of parameters. It is possible to switch among these set of parameters
by clicking on the Marker Out 1, …, the Marker Out 4 icons located in the
Marker Out Main Setting Bar.
The available parameters for the markers out are:
Marker Out Setting
Description
Voltage Level [V]
it sets the Marker high level Voltage. The Low level is fixed to 0V.
The Voltage Level range is 1 V to 2.5 V on 50 Ohm load (twice on open
load).