Operator’s Manual Simple AFG Application
21
Marker Out Specification Value
Connector
1 BNC for each pair of channels on the Front Panel
Output impedance
50 Ω
Output level (into 50 Ω)
1 V to 2.5 V
Important Note
:
the Marker Out 1 is linked to the Channel 1 and Channel 2, the Marker Out 2 is linked to the Channel
Out 3 and Channel Out 4, the Marker Out 3 is linked to the Channel Out 5 and Channel Out 6, the
Marker Out 4 is linked to the Channel Out 7 and Channel Out 8.
Model
Marker Out Connectors
T3AWG3352/3252
1 BNC on the Front Panel
T3AWG3354/3254
2 BNCs on the Front Panel
T3AWG3358/3258
4 BNCs on the Front Panel
Trigger In
The Trigger In (Trigger In connector on the front panel) allows to control the generation by an external
signal source. It has a selectable impedance of 1 kOhm or 50 Ohm. In Continuous mode the Trigger In
doesn’t have any effect.
Trigger In Specification Value
Connector
BNC on the Front Panel
Number of connectors
1
Input impedance
1
kΩ or 50 Ohm selectable
Slope/Polarity
Positive or negative selectable
Trigger In signal (blue, top) that starts a burst of sine waveform (red, bottom)