RiverPro and RioPro ADCP Guide
October 2017
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Page 129
Vertical Beam Profile Commands
The following commands define the criteria used to collect the vertical beam profile data.
Vertical beam profiling extends the ping time approximately 25% for manual modes. Ping
time depends on water depth and number of bins.
RioPro systems do not include a vertical beam.
Standard Vertical Beam Profile Commands
This section lists the vertical beam profile commands.
>Z?
Available Commands:
ZB 0 -------------------- Bandwidth [0=Wide (25%), 1=Narrow (6.25%)]
ZC 064 ------------------ Correlation Threshold (counts) [0..255]
ZD 111100000 ------------ Data Out {v;c;a;p;s;*;*;*;*}
ZF 020 ------------------ Blanking Distance (cm) [0-500]
ZG 1 -------------------- Gain [0=low, 1=high]
ZM 2 -------------------- V-Beam Profile Mode [2=Linear, 12=LinPhasePlaneAvg]
ZN 010 ------------------ Number of Bins [1-200]
ZO 004 ------------------ Number of Mode-12 Subpings [0-20]
ZP 000 ------------------ Number of Pings [0-999]
ZS 010 ------------------ Bin Size (cm) [2-500]
ZV 250 ------------------ Ambiguity Velocity (cm/s) [5-700]
Z? ---------------------- Display Z-Command Menu
>
ZB – Vertical Beam Bandwidth
Purpose:
Sets the vertical beam profile bandwidth (sampling rate).
Format:
ZBn
Range:
n = 0=Wide (25%), 1=Narrow (6.25%)
Default:
ZB0
Recommended Setting.
The default setting for this command is recommended for most applications.
Description:
Narrow bandwidths allow the RiverPro to profile farther, but the standard deviation is
increased by as much as 2.5 times.
ZC – Vertical Beam Correlation Threshold
Purpose:
Sets the minimum threshold for correlation magnitude that vertical beam profile velocity
data must meet to be considered valid.
Format:
ZCnnn
Range:
nnn = 0 to 255 counts
Default:
ZC064
Recommended Setting.
The default setting for this command is recommended for most applications.