RiverPro and RioPro ADCP Guide
October 2017
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Page 109
PC4 – Display Voltage Monitor ADC Data
The PC4 Batt (V) value is ~ 0.8 volts below the supply voltage. VDD1 and VDD3 should be within 0.01
volts of the values shown in the example data.
>pc4
Battery ADC data is sampled and displayed in a loop.
Press any key to exit the loop.
Count Chan0 Chan1 Chan2 Chan3 Batt(V) Batt(A) VDD1 VDD3
\ 11 0685 00c6 0d49 0b7f 11.878 0.121 3.328 1.797
PC40 – Display Scrolling Voltage Monitor ADC Data
PC40 is the same as PC4 except the data is displayed in a repeating loop at approximately 0.5-sec update
rate.
>pc40
Battery ADC data is sampled and displayed in a loop.
Press any key to exit the loop.
Count Chan0 Chan1 Chan2 Chan3 Batt(V) Batt(A) VDD1 VDD3
0 0684 00ca 0d49 0b80 11.871 0.123 3.328 1.797
0 0683 00e2 0d49 0b7f 11.865 0.138 3.328 1.797
0 0684 00ba 0d49 0b80 11.871 0.114 3.328 1.797
0 0684 00b2 0d49 0b7f 11.871 0.109 3.328 1.797
>
PC5 – Display BlueTooth RSSI Data
TRDI use only.
PC50 – Display Scrolling BlueTooth RSSI Data
TRDI use only.
PC6 – Display GPS RSSI Data
TRDI use only.
PC60 – Display Scrolling GPS RSSI Data
TRDI use only.