APPENDIX
211
The transition filter parameters detect changes in the specified
condition register bits (numeric suffixes 1 to 16) and overwrite the
extended event register in the following ways.
Condition
Definition
RISE
The specified extended event register bit is set to 1 when the
corresponding condition register bit changes from 0 to 1.
FALL
The specified extended event register bit is set to 1 when the
corresponding condition register bit changes from 1 to 0.
BOTH
The specified extended event register bit is set to 1 when the
corresponding condition register bit changes from 0 to 1 or
from 1 to 0.
NEVer
Always zero.
The following table describes the Standard Event Register
Bit
Name
Decimal
Definition
0
Operation
Complete
1
All commands prior to and including
*OPC have been executed.
1
Not Used
2
(Reserved for future use)
2
Query Error
4
The instrument tried to read the output
buffer but it was empty. Or, a new
command line was received before a
previous query has been read. Or, both the
input and output buffers are full.
3
Device Error
8
A device error, including a self-test error or
calibration error, occurred (an error in the -
300 range or any positive error has been
generated).
4
Execution
Error
16
An execution error occurred (an error in
the -200 range has been generated).
5
Command
Error
32
A command syntax error occurred (an
error in the -100 range has been
generated).
6
Not Used
64
(Reserved for future use)
7
Power On
128
Power has been cycled since the last time
the event register was read or cleared.
Содержание Everywhereyoulook T3PM1100
Страница 1: ...T3PM1100 Digital Power Meter User Manual ...
Страница 75: ...MEASUREMENT AND OTHER FUNCTIONS 75 3 Press ESC button to return back to original display mode ESC ...
Страница 119: ...COMMAND OVERVIEW 119 CR LF The most common EOL character is CR LF Message Separator EOL or semicolon Command Separator ...
Страница 191: ...COMMAND OVERVIEW 191 Example STATUS QMESSAGE ON STATUS QMESSAGE STATUS QMESSAGE 1 ...
Страница 209: ...APPENDIX 209 Status system The diagram below is a description of the status system ...
Страница 213: ...APPENDIX 213 Dimensions Units mm ...