COMMAND OVERVIEW
135
Example
:COMMUNICATE:REMOTE ON
:COMMUNICATE:REMOTE?
->:COMMUNICATE:REMOTE 1
:COMMunicate:STATus
Description
Returns and clears the line-specific status.(Only
for RS-232)
Query Syntax
:COMMunicate:STATus?
Return parameter
(each status bit)
Bit 0
Bit 1
Parity error.
Framing error.
Bit 2
Bit 3 and
higher
Noise error Break character detection.
Always zero.
Example
:COMMUNICATE:STATUS?
->0
Note
• When an event occurs, the corresponding bit is set
in the status.
• When the bit is read, it is cleared.
• Zero is returned for interfaces other than RS-232.
:COMMunicate:VERBose
Description
Sets or returns whether the response to a query is
returned fully spelled out or in its abbreviated
form.
Syntax
:COMMunicate:VERBose {<Boolean>|OFF | ON}
Query Syntax
:COMMunicate:VERBose?
Parameter
<Boolean>0 OFF
<Boolean>1 ON
Return parameter 0
Turn the verbose function off.
Query
Set
Query