COMMAND OVERVIEW
153
Example
:INPUT:CURRENT:SRATIO:ELEMENT11 10
:INPUT:CURRENT:SRATIO:ELEMENT11?
->:INPUT:CURRENT:SRATIO:ELEMENT1 EXT1,10.000
[:INPut]:RCONfig
Description
Sets or returns the on/off state of the range
configuration (valid range selection) feature.
Syntax
[:INPut]:RCONfig {<Boolean>|OFF|ON}
Query Syntax
[:INPut]:RCONfig?
Parameter
<Boolean>0 OFF
<Boolean>1 ON
Return parameter 0
Turn the range configuration feature off.
1
Turn the range configuration feature on.
Example
:INPUT:RCONFIG ON
:INPUT:RCONFIG?
->:INPUT:RCONFIG 1
[:INPut]:SCALing
Description
Returns all scaling settings.
Query Syntax
[:INPut]:SCALing?
Return parameter <String>
[:INPut]:SCALing[:STATe]
Description
Sets or returns the scaling on/off state.
Syntax
[:INPut]:SCALing[:STATe] {<Boolean>|OFF|ON}
Query Syntax
[:INPut]:SCALing[:STATe]?
Parameter
<Boolean>0 OFF
<Boolean>1 ON
Set
Query
Query
Set
Query