APPENDIX
199
Specifications
Below are the basic conditions required to operate the T3PM1100
within specification:
Calibration: Yearly
Operating Environment:
18~28 ˚C (64.4~82.4˚F)
Humidity: <80%RH,
Accuracy: ± (% of r % of range)
The specifications apply when the unit is warmed up for at least
30 minutes and operates in the slow rate.
The power supply cable must be grounded to ensure accuracy.
Input voltage and current must be standard sine wave.
The power factor must be 1.
The crest factor must be 3.
The common-mode voltage must be zero.
General Specifications
Specification Conditions:
Temperature: 23
˚
C±5
˚
C
Humidity: <80%RH(non-condensing)
Operating Environment: (0~40
˚
C)
Temperature Range: 30~40
˚
C, Relative Humidity: <70%RH(non-condensing);
>40
˚
C, Relative Humidity: <50%RH(non-condensing)
Indoor use only
Altitude: <2000 meters
Pollution degree 2
Storage Conditions (-40~70
˚
C)
Humidity: <90%RH(non-condensing)
General:
Power Source: 100-240 VAC 50/60Hz
Power Consumption: Max 30VA
Bench Dimensions: 268 mm (W) X 107 mm (H) X 379 mm (D) (w/t bumpers)
Weight: Approximately 2.9 kg
Содержание Everywhereyoulook T3PM1100
Страница 1: ...T3PM1100 Digital Power Meter User Manual ...
Страница 75: ...MEASUREMENT AND OTHER FUNCTIONS 75 3 Press ESC button to return back to original display mode ESC ...
Страница 119: ...COMMAND OVERVIEW 119 CR LF The most common EOL character is CR LF Message Separator EOL or semicolon Command Separator ...
Страница 191: ...COMMAND OVERVIEW 191 Example STATUS QMESSAGE ON STATUS QMESSAGE STATUS QMESSAGE 1 ...
Страница 209: ...APPENDIX 209 Status system The diagram below is a description of the status system ...
Страница 213: ...APPENDIX 213 Dimensions Units mm ...