COMMAND OVERVIEW
159
:INTEGrate:FUNCtion
Description
Sets or returns the integration function.
Syntax
:INTEGrate:FUNCtion {WATT|AMPEre}
Query Syntax
:INTEGrate:FUNCtion?
Parameter/
Return parameter
WATT
Select the integration function watt.
AMPEre
Select the integration function ampere.
Example
:INTEGRATE:FUNCTION WATT
:INTEGRATE:FUNCTION?
->:INTEGRATE:FUNCTION WATT
:INTEGrate:TIMer
Description
Sets or returns the integration timer value.
Syntax
:INTEGrate:TIMer {<NRf>,<NRf>,<NRf>}
Query Syntax
:INTEGrate:TIMer?
Parameter/
Return parameter
1st <NRf>
0 to 9999 (hours)
2nd <NRf>
0 to 59 (minutes)
3rd <NRf>
0 to 59 (seconds)
Example
:INTEGRATE:TIMER 1,0,0
:INTEGRATE:TIMER?
->:INTEGRATE:TIMER 1,0,0
:INTEGrate:STARt
Description
Starts integration.
Syntax
:INTEGrate:STARt
Example
:INTEGRATE:START
Set
Query
Set
Query
Set